Atec Agilent-E4980A User Manual
Page 12
Advertising

12
Effect by impedance of DUT
Table 14. For impedance of DUT below 30 Ω, the following value is added.
Test
Impedance of DUT
frequency [Hz]
1.08 Ω ≤ |Zx| < 30 Ω
|Zx| < 1.08 Ω
20 - 1 M
0.05%
0.10%
1 M - 2 M
0.10%
0.20%
Table 15. For impedance of DUT over 9.2 k Ω, the following value is added.
Test
Impedance of DUT
frequency [Hz]
9.2 kΩ < |Zx| ≤ 92 kΩ
92 kΩ < |Zx|
10 k - 100 k
0%
0.05%
100 k - 1 M
0.05%
0.05%
1 M - 2 M
0.10%
0.10%
Effect of cable extension
When the cable is extended, the following element is added per one meter.
0.015 % × (Fm/1 MHz)
2
× (L_cable)
2
Fm [Hz]
Test Frequency
L_cable [m]
Cable length
Advertising
See also other documents in the category Atec Equipment:
- Anritsu-PIM-MW82119A (2 pages)
- Boonton-PIM31 (6 pages)
- AWT-PIM-S1L-Tetra Series (2 pages)
- AWT-PIM-S1P Series (2 pages)
- AWT-PIM-S1L Series (2 pages)
- AWT-PIM-Expandable Series (2 pages)
- AWT-PIM-Single Series (2 pages)
- CCI-PimPro (4 pages)
- JDSU-FST-2310 (10 pages)
- JDSU-T-Berd-6000 (8 pages)
- Agilent-N9912A (2 pages)
- Agilent-E5515C (4 pages)
- Agilent-E4406A (20 pages)
- Agilent-N4010A (16 pages)
- Anritsu-S412E (19 pages)
- Anritsu-S810D-S820D (2 pages)
- Anritsu-S820E (16 pages)
- Anritsu-MT8221B (28 pages)
- Anritsu-MT8221B (24 pages)
- Anritsu-S412D (16 pages)
- Anritsu-MT8222A (8 pages)
- Anritsu-MT8220T (28 pages)
- Anritsu-MT8212E-MT8213E (32 pages)
- Anritsu-S332D-31D (12 pages)
- Anritsu-MT8212A (2 pages)
- Bird-SA Series (2 pages)
- Anritsu-S331E-S332E-S361E-S362E (16 pages)
- Anritsu-S331L (12 pages)
- Advantest-Q8163 (1 page)
- Agilent-83557A-83558A (4 pages)
- Agilent-11896A (5 pages)
- Agilent-8169A (6 pages)
- Agilent-81689A_B-81649A (6 pages)
- Agilent-8163A-81634A (8 pages)
- Agilent-81624B (11 pages)
- Agilent-81618A (112 pages)
- Agilent-8703A (16 pages)
- Agilent-8156A (8 pages)
- Advanced-Fiber-Solutions-OLK51 Series (5 pages)
- Anritsu_MS9720A (12 pages)
- Anritsu-ML9001A (4 pages)
- Corning-Optivisor400 (6 pages)
- Anritsu_CMA5000a (4 pages)
- Agilent-E6008B (12 pages)
- AFL-Noyes-FLX3 (6 pages)