Atec Agilent-E4980A User Manual
Page 23

23
DC resistance (Rdc) accuracy
Absolute measurement accuracy Aa
Absolute measurement accuracy Aa is given as
Equation 15.
Aa = Ae + Acal
Aa
Absolute accuracy (% of reading value)
Ae
Relative accuracy (% of reading value)
Acal Calibration accuracy
Relative measurement accuracy Ae
Relative measurement accuracy Ae is given as
Equation 16.
Ae = [Ab + (Rs /|Rm|+ Go × |Rm|) Ч 100 ] Ч Kt
Rm Measurement
value
Ab Basic
accuracy
Rs
Short offset [Ω]
Go Open
offset
[S]
Kt
Temperature coefficient
Calibration accuracy Acal
Calibration accuracy Acal is 0.03%.
Basic accuracy Ab
Table 42. Basic accuracy Ab is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
1.00%
2.00%
MED
0.30%
0.60%
Open offset Go
Table 43. Open offset Go is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
50 nS
500 nS
MED
10 nS
100 nS
Short offset Rs
Table 44. Short offset Rs is given below.
Measurement
Test signal voltage
time mode
≤ 2 Vrms
> 2 Vrms
SHORT
25 mΩ
250 mΩ
MED
5 mΩ
50 mΩ