Test signal – Atec Agilent-E4980A User Manual
Page 4

Table 1. Trigger delay time
Range
0 s - 999 s
Resolution
100 µs (0 s - 100 s)
1 ms (100 s - 999 s)
Table 2. Step delay time
Range
0 s - 999 s
Resolution
100 µs (0 s - 100 s)
1 ms (100 s - 999 s)
Measurement terminal: Four-terminal pair
Test cable length: 0 m, 1 m, 2 m, 4 m
Measurement time modes: Short mode, medium mode, long mode.
Table 3. Averaging
Range
1 - 256 measurements
Resolution
1
Test signal
Table 4. Test frequencies
Test frequencies
20 Hz - 2 MHz
Resolution
0.01 Hz (20 Hz - 99.99 Hz)
0.1 Hz (100 Hz - 999.9 Hz)
1 Hz (1 kHz - 9.999 kHz)
10 Hz (10 kHz - 99.99 kHz)
100 Hz (100 kHz - 999.9 kHz)
1 kHz (1 MHz - 2 MHz)
Measurement accuracy
± 0.01%
Table 5. Test signal modes
Normal
Program selected voltage or current at the measurement
terminals when they are opened or short-circuited, respectively.
Constant
Maintains selected voltage or current at the device under test
(DUT) independently of changes in impedance of DUT.
4