Effect of cable length (Short offset)

Table 45. The following value is added to Rs when the cable is extended.

Cable length
1 m

2 m

4 m

0.25 mΩ

0.5 mΩ

1 mΩ

Temperature coefficient Kt

Table 46. Temperature coefficient Kt is given below.

Temperature [°C]

Kt

0 - 18

4

18 - 28

1

28 - 55

4

Other options

Option 002 (Bias current interface): Adds a digital interface to allow the E4980A LCR
meter to control the Agilent 42841A bias current source.

Option 005 (Entry model): Economy option with less measurement speed.
Same measurement accuracy as the standard model.

Option 007 (Standard model): Upgrade to the standard model.

Option 201 (Handler interface): Adds handler interface.

Option 301 (Scanner interface): Adds scanner interface.

Note

Option 007 can be installed only in the E4980A