Fm scale dac offset calibration, Description, Required test equipment – Atec Agilent-8563E User Manual
Page 110: Procedure
Advertising

ESG Family Signal Generators
Adjustments
FM Scale DAC Offset Calibration
Calibration Guide
5-13
FM Scale DAC Offset Calibration
Description
This test is used to remove the offset associated with the FM SCALE DAC operational
amplifier located on the reference board. This calibration results in a DAC value for FM
OFFSET DAC 2. After this DAC value has been properly adjusted, the effects of the FM
SCALE DAC value on the offset will be minimized.
Required Test Equipment
• HP/Agilent 53132A Option 050 Universal Counter
Procedure
Figure 5-10 FM Scale DAC Offset Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.