FM/PM YO Frequency Compensation Calibration (ESG-AP and ESG-DP only)
Calibration Guide
5-23
FM/PM YO Frequency Compensation Calibration
(ESG-AP and ESG-DP only)
Description
This adjustment calibrates the FM/PM YIG oscillator frequency compensation latches on
the YO driver assembly. These latches adjust the flatness of the YO FM/PM paths for rates
greater than 100 kHz. The
“FM In-Band DAC Offset Calibration”
“FM/PM Out-of-Band Calibration (ESG-AP and ESG-DP only)”
on page 5-22 must be
performed prior to this adjustment.
Required Test Equipment
• HP/Agilent 8563E Spectrum Analyzer
• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator
Procedure
Figure 5-19 FM/PM YO Freq. Compensation Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.