Baseband generator i/q gain and offset adjustment, Description, Required test equipment – Atec Agilent-8563E User Manual
Page 140: Procedure

ESG Family Signal Generators
Adjustments
Baseband Generator I/Q Gain and Offset Adjustment (ESG-A and ESG-D Options UN3 and UN4 only)
Calibration Guide
5-43
Baseband Generator I/Q Gain and Offset Adjustment
(ESG-A and ESG-D Options UN3 and UN4 only)
Description
This adjustment uses a DVM to set the baseband generator’s four potentiometers (I Gain, I
Offset, Q Gain, and Q Offset) to minimize the dc offset and set the AC voltage level to
0.5 V
pk
(into 50 ohms) at the rear-panel I and Q output connectors.
Required Test Equipment
• HP/Agilent 3458A Digital Multimeter
• HP/Agilent 10100C 50
Ω
Feed Through
Procedure
Figure 5-35
Baseband Generator I/Q Gain and Offset Adjustment Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display. See