Table 4-2, Table 4-3 – Atec Agilent-8563E User Manual
Page 97

Performance Tests
ESG Family Signal Generators
Performance Test Records
4-48
Calibration Guide
Table 4-2 ESG Family Signal Generators Timebase Aging Rate
Performance Test Record
Table 4-3
ESG-AP and ESG-DP Series Signal Generators Phase Noise and
Residual FM Performance Test Record
Model _____________
Report Number ____________________________
Date ___________
Test Description
Results
Specification
Measurement
Uncertainty
Timebase Aging Rate
ESG-AP, ESG-DP, or
Option 1E5 only
_____________________
<
±
0.0005 ppm/day
5.6
×
10
-11
per day
Model
___________________
Serial No.
__________________
Date
___________________
Tester
____________________
Phase Noise (20kHz Offset)
Frequency (GHz)
Specification
(dBc/Hz)
Measurement
Pass/Fail?
0.5
< –134
1.0
< –130
2.0
< –123
3.0
< –120
4.0
< –118
Residual FM (CCITT Weighting)
Frequency (GHz)
Specification (Hz)
Measurement
Pass/Fail?
0.5
< 0.5
1.0
< 1
2.0
< 2
3.0
< 3
4.0
< 4