Test set-up, Test set-up example, Table top equipment – Atec Noiseken-ESS-200AX User Manual

Page 11: Electrostatic discharge simulator

Advertising
background image

Electrostatic Discharge Simulator

9

0.1m

ESD simulator

Isolation tranformer

ESD simulator

Insulation pallet
(h=0.1m)

Ground reference plane

470kΩ resistor

Vertical Coupling Plane
(0.5m x 0.5m)

Typical position for
direct application

Typical position for
indirect discharge to VCP

EUT

0.1m

EUT power

(AC)

ESD simulator

power

>1m distance between
the EUT and laboratory walls

Protective earth

Earth cable

Vertical Coupling Plane
(0.5m x 0.5m)

Typical position for
indirect discharge to VCP

>1m distance between
the EUT and laboratory walls

Typical position for
indirect discharge to HCP

Isolation tranformer

ESD simulator

Horizontal coupling plane

Ground reference plane

Wooden table

Insulation sheet

0.1m

Typical position for
direct application

EUT power

(AC)

ESD simulator power

EUT

0.1m

0.8m

470kΩ resistor

Protective earth

Earth cable

IEC 61000-4-2 standard

TEST SET-UP

Ground reference plane: A copper or aluminum sheet of
0.25 mm minimum thickness: other materials may be used
but they shall have at least 0.65 mm minimum thickness.
The minimum size is 1㎡ . The exact size depends on the
EUT. It shall project beyond the EUT or coupling plane by
at least 0.5 m on all sides. It shall be connected to the pro-
tective earth
Coupling planes: These planes shall be constructed from
the same material and thickness as that of the ground refer-
ence plane and shall be connected to the ground reference
plane via a cable with a 470k ohm resistor located at each
end.

Test set-up for test performed in laboratories:
A ground reference plane shall be provided on the fl oor of
the laboratory.
The EUT shall be connected to the grounding system
and arranged and connected according to its installation
specifi cations. A distance of 1 m minimum shall be provided
between the EUT and any metallic structure.
The discharge return cable of the test generator shall
be connected to the ground reference plane, and this
connection shall be of low impedance.
In cases where the length of the cable exceeds the length
necessary to apply the discharges to the selected points,
the excess length shall be placed non-inductively off the
ground reference plane and shall not come closer than 0.2
m to other conductive parts in the test set-up.

TEST SET-UP EXAMPLE

Table top equipment

Test set-up for table-top equipment, laboratory
tests
A wooden table of 0.8m height shall be set on
the ground plane. 1.6m x 0.8 m horizontal and
0.5m x 0.5 m vertical coupling planes shall be
put on the table. An insulating support of 0.5
mm thickness shall be inserted between the
EUT/cables and the horizontal coupling plane.

An insulation support of 0.1m thickness
shall be used. 0.5m x 0.5m vertical coupling
plane shall be used for indirect application of
discharges.

Floor-standing equipment

Test set-up for table-top equipment, laboratpry tests