Electrostatic Discharge Simulator

5

ESD Simulator ESS-2002EX

&

&

TC-815R

ESD Simulator

Conforming to IEC61000-4-2

A manual ESD simulator that is both lightweight and
compact. A completely new design has made the product
easier to use, safer and more reliable.

FEATURES

Easy to use self-explanatory control panel
Preset discharge interval and count
Preset IEC standard test levels
“V STEP” function enables the voltage setting can be altered
in steps of the desired preset increment (or decrement)

CONTROLS INDICATORS AND DISPLAYS

Specifi cations

Item

Specifi cations

Output polarity

Positive or negative

Output voltage

0.20kV ∼ 30.0kV± 5% (30.5kV max)

0.20kV ∼ 1.0kV: 0.05kV step, 1.0kV ∼ 30.0kV:0.1kV step

Discharge interval
(contact discharge)

Normal mode: 0.05s ∼ 9.99s± 10%

Extra mode : 1.00s ∼ 9.99s± 10%

0.01s step

No. of times of discharge

1 ∼ 999 times preset on the coutner or continuous

Discharge mode

Air/contact discharge

Radiation level

Normal/Extra mode

Trigger mode

Gun trigger or main unit trigger

Preset IEC levels

Contact discharge: 2.0kV, 4.0kV, 6.0kV, 8.0kV (4 test voltages preset)
Air discharge: 2.0kV, 4.0kV, 8.0kV, 15.0kV (4 test votlages preset)

Preset step voltage mode
(“V STEP” function)

Step setting: 0.00kV ∼ 10kV

0.2kV ∼ 1.0kV: 0.05kV step, 1.0kV ∼ 30.0kV: 0.1kV step

Each time Increment or Decrement key is pressed, the test voltage changes by the preset step

Warning lamp

Lights up while the HV circuitry is on.
Blinks while discharges are in progress

Discharge gun

TC-815R

Charging resistance

10MΩ(53 MΩ for combination with TC-815R discharge gun)

Power supply

100 ∼ 240V AC 50/60 Hz ± 10%

Power consumption

62VA

Operating temperature and humidity

+ 15 ∼ +35℃ 25 ∼ 75% (no condensatoin)

Dimensions and Weight

340 (W) X 200 (H) X 170 (D) mm (projection excluded) Approx 4.6 kg

+/- Selects the polarity

Displays output voltage, discharge interval and number of discharges settings.

Increment and decrement keys: change the value for the selected parameter

CONTACT/AIR selects either air or contact discharge

V STEP: when this enabled, each time Increment or Decrement key is pressed, the test voltage changes by
the preset step.

TRIGGER: selects either CONTROLER (TRIGGER key on ESS main unit) or TC-815R gun as the operative trigger

Warning lamp: alight while the HV circuitry is on and blinks while discharges are in progress

STOP key: turns off the HV circuitry

START key: turns on the HV circuitry

SELECT key: selects the parameter to be changed. Each time the key is pressed, VOLTAGE, INTERVAL and COUNT
indicator starts blinking in turn. While blinking, the value can be changed by using the Increment or Decrement key

IEC LEVEL: selects either of the preset IEC specified test voltages. The preset voltages are 2.0kV, 4kV, 6kV
and 8kV for contact discharge and 2.0kV, 4.0kV, 8.0kV and 15.0kV for air discharge as per IEC61000-4-2

EXTRA/NORMAL: provides the selection of NORMAL or EXTRA in terms of the strength of radiation from the
discharge gun. The EXTRA mode reduces the radiation. In the EXTRA mode, faster than 1.0s setting for
discharge interval is not accepted.

ESS-2002EX & TC-815R (with PS-806 gun stand)