Constant resistance mode, Timed operation – B&K Precision 8500-8526 - Manual User Manual

Page 33

Advertising
background image

In constant power mode, the DC load will cause a constant power to be dissipated in the load. To
set up the DC Load to operate in constant power mode and in the steady state condition, use the
following keystrokes:

Key

Display

P-set

If you press this key after powering up and there is no currently-
stored constant power value, you'll be prompted for the desired
power level. If there was a power value already stored, it will be
momentarily displayed and the instrument will be in CW mode. If
you wish to change the power setting, press P-set again. The

OFF

annunciator will be displayed.

On/Off

Turns the constant power load on. The

CW

annunciator will be on.

or

Momentarily display the power level and the set power value.

On/Off

Turns the load off.

To operate the DC Load in CW mode in a transient condition, see the Transient operation section.
To operate the DC Load in CW mode in a dynamic condition, see the Lists section.

Constant resistance mode

In constant resistance mode, the DC load will behave as a fixed resistance value. To set up the DC
Load
to operate in constant resistance mode and in the steady state condition, use the following
keystrokes:

Key

Display

R-set

If you press this key after powering up and there is no currently-
stored constant resistance value, you'll be prompted for the desired
resistance. If there was a resistance value already stored, it will be
momentarily displayed and the instrument will be in CR mode. If
you wish to change the resistance setting, press R-set again. The

OFF

annunciator will be displayed.

On/Off

Turns the constant resistance load on. The

CR

annunciator will be

on.

or

Momentarily display the power level and the set resistance value.

On/Off

Turns the load off.

To operate the DC Load in CR mode in a transient condition, see the Transient operation section.
To operate the DC Load in CR mode in a dynamic condition, see the Lists section.

Timed operation

The DC Load can be set to allow timed operation. When you turn the load on, it stays on for the
specified time, then turns itself off.
To set up timed operation, use the following keystrokes:

8500 DC Load Series

Version: 030614

Page 33 of 77