0x50 set timer value of for load on, 0x51 read timer value for load on, 0x52 disable/enable timer for load on – B&K Precision 8500-8526 - Manual User Manual

Page 66: 0x53 read timer state for load on, 0x54 set communication address, 0x55 enable/disable local control

Advertising
background image

Byte

offset

Meaning

7-24

Reserved

0x50 Set timer value of for LOAD ON

Byte

offset

Meaning

3 to 4

2 byte little-endian integer specifying the time in units of 1 second

5-24

Reserved

0x51 Read timer value for LOAD ON

Byte

offset

Meaning

3 to 4

2 byte little-endian integer specifying the time in units of 1 second

5-24

Reserved

0x52 Disable/enable timer for LOAD ON

Byte

offset

Meaning

3

0 is disable timer
1 is enable timer

4-24

Reserved

0x53 Read timer state for LOAD ON

Byte

offset

Meaning

3

0 is disable timer
1 is enable timer

4-24

Reserved

0x54 Set communication address

Byte

offset

Meaning

3

2 byte little-endian integer specifying the address. Must be between 0 and
0xFE, inclusive.

4-24

Reserved

0x55 Enable/disable LOCAL control

Byte

offset

Meaning

3

0 means to disable the Local key on the front panel

8500 DC Load Series

Version: 030614

Page 66 of 77