Transducer Positioning

1. Place the upstream transducer in position and secure with a

mounting strap. Place straps in the arched groove on the end of

the transducer. A screw is provided to help hold the transducer

onto the strap. Verify that the transducer is true to the pipe and

adjust as necessary. Tighten the transducer strap securely.

2. Place the downstream transducer on the pipe at the calculated

transducer spacing. See

Figure 27

. Apply firm hand pressure.

If signal quality is greater than 10, secure the transducer at this

location. If the signal quality is not 10 or greater, using firm hand

pressure slowly move the transducer both towards and away

from the upstream transducer while observing signal quality.

Signal quality can be viewed on the DXN’s display. Clamp the

transducer at the position where the highest signal quality

is observed. The factory default signal quality setting is 10%,

however there are many application specific conditions that may

prevent the signal quality from attaining this level. For the DXN,

signal levels less than 10% will not give readings.

NOTE:

N

Signal quality readings update only every few seconds, so it

is advisable to move the transducer 1⁄8", wait, see if signal is

increasing or decreasing and then repeat until the highest level

is achieved.

Figure 27: Transducer Positioning

3. If after adjustment of the transducers the signal quality does not rise to above 10%, then select an alternate

transducer mounting method. If the mounting method was W-Mount, then re-configure the transmitter for V-Mount,

move the downstream transducer to the new spacing distance.

NOTE:

N

As a rule, the DTTL transducers should be used on pipes 24" and larger. Do not use DTTL transducers for application

on a pipe smaller than 4". Consider application of the DTTL transducers on pipes smaller than 24" if there are less

quantifiable aspects such as - sludge, tuberculation, scale, rubber liners, plastic liners, thick mortar liners, gas bubbles,

suspended solids, emulsions, and smaller pipes that are perhaps partially buried where a V-Mount is required

or desired.

Transducer

Spacing

Site Name

0.00 GPM

1/16/2013

2:18 PM

Meter

Adv

Cal

System

Display

Site

Log

I/O

Create

Lookup

Liner

Doppler

Fluid

Pipe

Transit

Required Spacing (in)

Transducer

Transducer Mount

DTTN 1MHz

Z

12.14

12--1/8”

DXN Portable Ultrasonic Measurement System

Page 24

February 2013

Advertising