Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
63
FlexCor
™
Model CMF Series
Example; setting the
0-point
Keypad operation
Implementation
Display
Push for 3 sec.
To access the
Password
user password
0000
Push once
To unlock password
CHANGE
0000
Push once
To enter 1000 as
CHANGE
password
1000
Push once
To lock password and to
CONV.SETUP MODE>
enter the menu
Basic settings
Push 4 times
To go to reset
Reset
mode menu
mode
Push once
To enter the reset
Totalizer 1 reset
mode menu
xxx. G
Push twice
To go to zero
Zero adjust
adjust menu
Push once
To enter the zero
Zero adjust
adjust menu
Auto
Push twice
To go to zero
Zero adjust
adjust activation
+xxxxxx kg/h
Push once
To activate the 0-point
Zero adjust
adjustment routine
+xxxxxx kg/h
Transmitter preforms
Counting up from 0
0-point adjustment
to 100%
New 0-point is calculated
Zero adjust
and stored in SENSORPROM
â
+yyyyyy kg/h
Push twice/
Transmitter reverts to
hold 3 sec.
standard operation
Menu detail