Commissioning

FlexCOR

Model CMF Series

52

Fluid Components Intl

Document 06EN003327 Rev. -

7.2

Menu build-up

The menu structure of a specific type of transmitter is shown in a menu overview map.
Details of how a specific parameter is set is shown in a menu detail map for the specific parameter.

The menu structure is valid for the title and subtitle line only. The upper line is for primary readings

only and will always be active with either mass flow rate, volume flowrate, density, temperature,
totalizer 1 or totalizer 2.

The menu is built up in two parts. An operator menu and a setup menu.

Operator menu
The operator menu is for daily operation. The operator menu is customized in the operator menu

setup. The signal converter always starts in the operator menu no. 1. The page forward and page

backward keys are used to step through the operator menus.

Setup menu
The setup menu is for commissioning and service only.

The setup menu is accessed by pressing the top up key for 2 seconds. The setup menu will operate
in two modes:

-

View mode

-

Setup mode

View mode is a read only mode. The pre-selected settings can only be scanned.

Setup mode is a read and write mode. The pre-selected settings can be scanned and changed.

Access to the setup mode is protected with a password. The factory set password is 1000.

Access to a submenu in the set up menu is gained by the lock key. A short press on the top up key

will bring back the previous menu. A long press (2 sec.) on the top up key will exit the setup menu

and bring back the operator menu no. 1.

7.2.1 Password

The SETUP MENU can be operated in two different modes:

VIEW MODE (Read only)

CHANGE MODE (Read and write mode)

The view mode is always accessed by pressing the forward key when in the password menu.

Access to change mode is protected by a user code. The user code is factory set to 1000, but can
be changed to any value between 1000 and 9999 in the change password menu.

The factory setting of 1000 can be re-established as follows:
-

Switch off power suppply

-

Press the TOP UP key while switching on the power supply

The user code is reset to 1000.