APPENDIX D - CE CONFORMANCE
Models MT86, MT86HT
D - 2
Doc. No. 003162 Rev. F
INSTALLATION CONFORMITY CRITERIA
Grounding
All enclosures must be grounded to earth ground through a path of less than 1 ohm.
Interconnecting Cables
All interconnecting cables between the instrument's local enclosure, remote enclosure, power source and mounting
device shall be enclosed in metal conduit. AC power input cabling shall be enclosed separately in conduit before
entrance into the instrument and is not to be combined with switch or monitor output cabling.