APPENDIX D - MT91 DELTA R PARAMETERS

FLUID COMPONENTS INTL

5.1.3 Verify Setup Outputs

[1] = Channel 1

[2] = Channel 2

PortUnit[1] 0 = Flow 1 = Temperature
PortUnit[2] 0 = Flow 1 = Temperature
Out Head[1] Hexadecimal Notation
Out Head[2] Hexadecimal Notation
Zero[1]

Zero output setting at either zero (0) or minimum
flow. (e.g. 4mA represents either 0 or minimum
flow)

Zero[2]
Span[1]

High point of output signal.(e.g. 20mA represents
high point)

Span[2]

5.1.4 Verify Setup Communication

Comm Mode 0 = EIA-232C 1 = EIA-422

2 = EIA-485

Comm Rate 0 = 4800

1 = 9600

2 = 19.2K

Comm Prot 0 = Protocol 1 1 = Protocol 2 2 = Protocol 3
485 Addr

Explanation of Parameters and Code

5.1.1 Verify Setup Flow

Flow Units

Customer Specified Unit

Flow Heads Hexadecimal Notation
Area Unit

Customer Specified Unit

Area

Area of Customer Pipe or Duct

TC Ena

Temperature Compensation Enabled

Corr Ena

Correction Enabled

Offset Ena

Offset Enabled

Filter

Filter Strength (See Chapter 3)

Density

Standard Density of Customer Media in LBM per

cubic feet.

Flow Fac

Flow Factor

5.1.2 Verify Setup Temperature

Temp Type Units used during calibration.
Temp Heads Hexadecimal Notation

Appedix D. MT91 Delta R Parameters

For each instrument there are “MT91 Delta R Parameters” sheets and a 4-20mA look up table inserted in the rear
page protector. Appendix D explains the codes found on the Delta R Parameters sheets. A sample of both the Delta
R Parameter sheet and the 4-20mA look up table follow the explanations.

Above each column is a menu level code. When verifying parameters, access the menu level and then scroll down
through the menu until the value in question appears on the screen. For example, to view the parameter “dRmin 1”,
press HOME, 5, 2, 3, and 1. (Verify

Þ

Verify Cal

Þ

Flow

Þ

Limits). Scroll down by pressing ENTR until “dRmin 1”

appears on the screen.

The instrument uses internal units of Standard Feet per Second and degrees Fahrenheit. Therefore, all values on this
sheet are in Standard Feet per Second. Refer to "Verify Standard vs. Actual Conditions" in Chapter 5 to convert
actual readings to standard readings.

In menu levels 5.1.1 and 5.1.3 active sensing points are designated by hexadecimal notation. Use the following
diagram and table to decipher which sensing points are active. For example, the number 0003 in hexadecimal
notation means sensing points 1 and 2 are active.

Group

Code

Hex

#

Group

Code

Hex

#

'0000

0

1000

8

'0001

1

1001

9

'0010

2

1010

A

'0011

3

1011

B

'0100

4

1100

C

'0101

5

1101

D

'0110

6

1110

E

'0111

7

1111

F

8 6 F 1

1 0 0 0

0 1 1 0

1 1 1 1

0 0 0 1

G r o u p 4

This hexidecimal

notation is found on the

Delta R Parameter Sheet

and on the Display.

This code indicates that

sensing points

1,5,6,7,8,10,11, and 16

are active.

S e n s i n g

Point 1

0 = O f f
1 = O n
Group 1 represents sensing points 1,2,3, and 4.
Group 2 represents sensing points 5,6,7,and 8.
Group 3 represents sensing points 9,10,11, and 12.
Group 4 represents sensing points 13,14,15, and 16.