BODTrak™ II procedures

20

5.2 Simplified procedure

Required apparatus:

BODTrak II bottle

Thermometer

Blender (optional)

Graduated cylinder

Required reagents:

1 nutrient buffer pillow

Table 4 Simplified sample volumes

BOD range mg/L

Sample volume mL

0 to 35

420

0 to 70

355

0 to 350

160

0 to 700

95

1. Heat or cool the

sample to 19 to 21 ºC

(66 F to 70 ºF).

2. Homogenize the

sample in a blender if it

contains large settleable

or floatable solids.

3. Choose the correct

sample size for the

sample range (

Table 4

).

Measure the sample into a

graduated cylinder.