7 special considerations, 1 sample dilution, 2 sample seeding – Hach-Lange BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) User Manual

Page 33: 3 sample temperature, 4 toxic materials, Fer to

Advertising
background image

BODTrak™ II procedures

33

5.7 Special considerations

5.7.1 Sample dilution

Unknown sample BOD effluent is typically in the 0 to 70 mg/L range. Unknown

sample BOD influent is typically in the 0 to 700 mg/L range. When the oxygen

requirement of a sample is more than 700 mg/L, dilute the sample with high-quality

distilled or deionized water.

Calculate the results to include the additional dilution factor. Example: If the BOD of

the sample is 1000 mg/L, dilute the sample 1:1 with distilled or deionized water. The

estimated BOD is now 500 mg/L. Use the sample volume specified in the table for

the 0 to 700 mg/L range of the chosen procedure. Multiply the instrument reading

result by 2. If using the Hach Standard Method procedure, continue with remaining

calculations.

5.7.2 Sample seeding

Some types of BOD samples do not contain sufficient bacteria to oxidize the

organic matter in the sample. Many industrial wastes are of this type. Some sewage

treatment plant effluents are chlorinated and essentially sterile. A BOD test cannot

be done in the absence of viable bacteria. To test such samples, seed each bottle

from a source known to contain a viable bacterial population.

Settled domestic wastewater plant influent or primary clarifier effluent are the

preferred sources of seed for most samples. Mixed liquor or undisinfected effluent

can be used for seed, but it is recommended to include a nitrification inhibitor.

Commercial seed sources are sometimes suitable. To prepare, see the instructions

from the manufacturer.

5.7.3 Sample temperature

Standard Methods for the Examination of Water and Wastewater, 21st Ed., 2005,

5210 D recommends an incubation temperature of 20 ±1 °C (68 °F) for the BOD

test. Put the BODTrak II instrument in an incubator that is adjusted to 20 ±1 °C. An

applicable BOD incubator is available from Hach (

section 8.1 on page 41

). Warm or

cool samples to 20 ± 1 °C.

Note: Instrument performance has not been validated at temperatures other than 20 °C.

5.7.4 Toxic materials

Industrial and chlorinated samples often contain toxic substances and require

special considerations when running BOD tests. Toxic materials in the sample will

cause decreased BOD values. Dilute the sample to minimize the toxic materials or

their effects. Refer to Standard Methods for the Examination of Water and

Wastewater, 21st edition, 5210 D.

Advertising