Section 1 specifications – Hach-Lange BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) User Manual

Page 5

Advertising
background image

5

Section 1

Specifications

Specifications are subject to change without notice.

Table 1 Specifications

General

Range

Selectable, 0 to 35, 0 to 70, 0 to 350, 0 to 700 mg/L

Dimensions

28.9 x 26 x 9.8 cm (11

3

/

8

x 10 ¼ x 3

7

/

8

inches)

External power supply

Input: 110 to 240 V, 50/60 Hz, Output: 24 V, UL CSA, and

TUV approved

Capacity

Six 492 mL bottles

Shipping weight

4 kg (8.8 lb)

Operating temperature

20 ºC (68 ºF)

Storage temperature

0 to 40 ºC (104 ºF)

Method performance specifications

Precision

On a standard containing 150 mg/L each of glucose and

glutamic acid, a single analyst using 6 BODTrak™ II

instruments and testing 44 samples obtained a mean of

235 mg/L BOD with a 95% confidence limit of distribution

of 11 mg/L or a range of

224 to 246 mg/L BOD.

Drift

Less than 3 mg/L BOD in 5 days

Resolution

1 mg/L BOD

Table 2 Certification

Certification

Hach Company certifies this instrument was tested thoroughly, inspected and found to meet

its published specifications when it was shipped from the factory. The BODTrak II has been

tested and is certified as indicated to the following instrumentation standards:
FCC Part 15, Sub-Part B, Class A Limits: Supporting test records by Intellistor, certified

compliance by Hach Company
Canadian Interference-Causing Equipment Regulation, ICES-003, Class A: Supporting

test records by Intellistor, certified compliance by Hach Company
EN 55011/CISPR 11(EMI) “B” Limits per 89/336/EEC EMC: Supporting test records by

Intellistor, certified compliance by Hach Company
EN 50082-1 (Immunity) per 89/336/EEC EMC: Supporting test records by Hach Company,

certified compliance by Hach Company. Standards include:

IEC 801-2 and EN 61000-4-2 (ESD)

IEC 801-3 and EN V50140 (RF & EM Field)

IEC 801-4 and EN 61000-4-4 (Fast Transient)

EN 61000-4-5 (Sur

ge)

Warranty: US 1 year; EU 2 year