Section 6 maintenance, 1 cleaning the instrument, 1 sample bottles – Hach-Lange BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) User Manual

Page 35: 2 stir bars and seal cups, 3 bottle fences, 2 stir bars and seal cups 6.1.3 bottle fences

Advertising
background image

35

Section 6

Maintenance

DANGER

Only qualified personnel should conduct the tasks described in this section

of the manual.

6.1 Cleaning the instrument

Clean spills on the BODTrak™ II instrument with a soft cloth which has been

dampened with deionized or distilled water.

6.1.1 Sample bottles

After each test, empty the sample bottles and flush them thoroughly with hot water.

Use a brush, hot water and soap to remove residue. Residue creates a BOD. Flush

the bottles thoroughly with tap water and finally with distilled or deionized water to

remove all detergent.

6.1.2 Stir bars and seal cups

Clean the stir bars with hot water and soap. Use a brush to remove deposits. Flush

with tap water and finally with distilled or deionized water to remove all detergent.

Carefully empty and rinse the seal cups with water. Invert to dry.

6.1.3 Bottle fences

The bottle fences prevent tipping of the bottles and provide tubing management

during storage. For storage, put the tubing into the opening in the bottle fence.

Wind the tubing counter-clockwise and secure the bottle cap inside the fence.

Advertising