Applications, Calibration, Reference standards – Hach-Lange REFO 3_REFO 3D User Manual

Page 12: Gloss measurement, C applications, C1 calibration, C2 reference standards, C3 gloss measurement

Advertising
background image

Operating Instructions REFO 3 / REFO 3-D

11

C Applications

C1 Calibration

It is recommended to calibrate at least once a day or after extended breaks. Before
starting to calibrate make sure that the standard is clean and the cloth removed.

The calibration is performed as follows:

•

Put REFO on the standard

•

Turn on REFO with key Meas/ON

•

Press key Meas/ON

•

During the calibration, the display reads *Cal* or M-Cal (for metal standard)

•

Wait for ok display

Calibration is only possible with one of programs P1-P3 set. The system recognizes
D- or M-mode automatically and shows it on the display before the standard number.

C2 Reference Standards

The chosen reference standard is standard A according to DIN 67530/ISO 2813 or
the optional metal standard according to ISO 7668 for high-gloss measurements.

C3 Gloss Measurement

•

Turn on REFO with key Meas/ON

•

Use the program key to select program P1

•

Use the angle selector key to select the viewing angle (the selected angle is

shown on the display)

•

Place REFO on the sample to measure

•

Press key Meas/ON

Now the measured value is shown on the display and stored automatically with con-
secutive numbering.