Exchanging the batteries, Entering standard number and calibration values – Hach-Lange REFO 3_REFO 3D User Manual

Page 8

Advertising
background image

Operating Instructions REFO 3 / REFO 3-D

7

Battery
Compartment

mains / charging
socket

Figure 3: REFO rear view

B2.2 Exchanging the batteries

Note: When exchanging the batteries, all stored measurings will be lost if the cover

of the battery compartment is lifted-off while the instrument is on. Therefore
wait for the automatical shut-down before exchanging the batteries.

To exchange the batteries:

•

Lift-off the cover of the battery compartment (see fig. 3)

•

Take-off all batteries

•

Put fresh batteries into the compartment as shown in the sketch

•

Snap the cover onto the battery compartment again.

•

B2.3 Entering standard number and calibration values

Standard number and calibration values need only be entered again if the standard is
changed. The 4-digit standard number and calibration values for 20°, 60° and 85°
are indicated on the rear of the standard they apply to.

1. Put REFO on the standard (remove cloth!)

2. Turn on REFO with key Meas/ON

3. Select program P3 with the program key