12

C4 Difference Gloss Measurement

Difference measurement means measuring against a reference. The measured dif-
ference between reference and sample is the indicated measured value.

•

Turn on REFO with key Meas/ON

•

Use the program key to select program P2

•

Use the angle selector key to select the viewing angle (the selected angle is in-

dicated on the display)

•

Place REFO on the reference

•

Press key Meas/ON to measure the reference. The reference measurement is

confirmed on the display with letter R next to the angle value

•

Place REFO on the sample to measure

•

Press key Meas/ON to measure the difference

Now the measured difference value is shown on the display and stored automatically
with consecutive numbering.

C5 Three-Angle Measurement

In three-angle measurements the sample is measured automatically at viewing
angles of 20°, 60° and 85°.

•

Turn on REFO with key Meas/ON

•

Use the program key to select program P3

•

Use the angle selector key to select the first viewing angle

•

Place REFO on the sample to measure

•

Press key Meas/ON

•

Wait for the measuring display

•

Call the other measurings with the angle selector key

Now the measured value is shown on the display. All of the three measurings are
stored together under a number. A maximum of 200 three-angle measurements can
be stored.