Error message

Displayed Error

Definition

Resolution

FLASH FAILURE

The external serial flash
memory Read/Write fails

Contact technical service

Event Log

Refer to

Table 3

for diagnostic device information.

Table 3 Event log

Event

Description

0: Power Up Event

Logs the power up time

1: Device Comm Loss

Reports the loss of communications
with a device. (Data: Device Index)

2: Device Comm Restore

Reports the restored communications
with a device (Data: Device Index)

3: Software Restart Event

Reports a software restart.

Classified errors and classified status

Table 4

,

Table 5

,

Table 6

,

Table 7

and

Table 8

show the classified error

register and the classified status 1-4 register flags for main
measurements. All the sensors and analyzers supply this signal quality
information at the same register address.

Table 4 Classified errors – register 49930

Bit

Error

Note

0

Measurement Calibration
Error

An error has occurred during the last
calibration.

1

Electronic Adjustment Error

An error has occurred during the last
electronic calibration.

2

Cleaning Error

The last cleaning cycle failed.

Table 4 Classified errors – register 49930 (continued)

Bit

Error

Note

3

Measuring Module Error

A failure has been detected in the
Measuring Module.

4

System Re-initialization
Error

Some settings have been detected
inconsistent and set to factory defaults.

5

Hardware Error

Any hardware error in general has been
detected.

6

Internal Communication
Error

A communication failure within the device
has been detected.

7

Humidity Error

Excessive humidity has been detected in
this device.

8

Temperature Error

Temperature within the device exceeds
specified limit.

9

Reserved for future use

Fixed at 0

10

Sample Warning

Some action is required with the sample
system.

11

Questionable Calibration
Warning

The last calibration was of questionable
accuracy.

12

Questionable Measurement
Warning

One or more of the measurements of the
device are of questionable accuracy
(Quality bad or out of range).

13

Safety Warning

A condition has been detected which may
result in a safety hazard.

14

Reagent Warning

Some action is required with the reagent
system.

15

Maintenance Required
Warning

Maintenance is required on this device.