Table 5 Classified status 1 – register 49931

Bit

Error

Note

0

Calibration in progress

The device has been placed in a
calibration mode. The measurements may
not be valid.

1

Cleaning in progress

The device has been placed in a cleaning
mode. The measurements may not be
valid.

2

Service/Maintenance menu

The device has been placed in a service or
maintenance mode in which the
measurements may not be valid.

3

Common error

Device recognised an error; see Error
Register for Error Class.

4

Measurement 0 Quality Bad

Precision of measurement is out of
specified limits.

5

Measurement 0 Low Limit

Measurement is below measurement
range.

6

Measurement 0 High Limit

Measurement is above measurement
range.

7

Measurement 1 Quality Bad

8

Measurement 1 Low Limit

9

Measurement 1 High Limit

10

Measurement 2 Quality Bad

11

Measurement 2 Low Limit

12

Measurement 2 High Limit

13

Measurement 3 Quality Bad

14

Measurement 3 Low Limit

15

Measurement 3 High Limit

Table 6 Classified status 2 – register 49932

Bit

Error

Note

0

Reserved for future use

Fixed at 0

1

2

3

4

Measurement 4 Quality Bad

Precision of measurement is out of
specified limits.

5

Measurement 4 Low Limit

Measurement is below measurement
range.

6

Measurement 4 High Limit

Measurement is above measurement
range.

7

Measurement 5 Quality Bad

8

Measurement 5 Low Limit

9

Measurement 5 High Limit

10

Measurement 6 Quality Bad

11

Measurement 6 Low Limit

12

Measurement 6 High Limit

13

Measurement 7 Quality Bad

14

Measurement 7 Low Limit

15

Measurement 7 High Limit

Table 7 Classified status 3 – register 49933

Bit

Error

Note

0

Reserved for future use

Fixed at 0

1

2

English 11