Table 8 – Hach-Lange SC 200 Modbus Module User Manual User Manual

Page 12

Advertising
background image

Table 7 Classified status 3 – register 49933 (continued)

Bit

Error

Note

3

4

Measurement 8 Quality Bad

Precision of measurement is out of
specified limits.

5

Measurement 8 Low Limit

Measurement is below measurement
range.

6

Measurement 8 High Limit

Measurement is above measurement
range.

7

Measurement 9 Quality Bad

8

Measurement 9 Low Limit

9

Measurement 9 High Limit

10

Measurement 10 Quality
Bad

11

Measurement 10 Low Limit

12

Measurement 10 High Limit

13

Measurement 11 Quality
Bad

14

Measurement 11 Low Limit

15

Measurement 11 High Limit

Table 8 Classified status 4 – register 49934

Bit

Error

Note

0

Reserved for future use

Fixed at 0

1

2

3

Table 8 Classified status 4 – register 49934 (continued)

Bit

Error

Note

4

Measurement 12 Quality
Bad

Precision of measurement is out of
specified limits.

5

Measurement 12 Low Limit

Measurement is below measurement
range.

6

Measurement 12 High Limit

Measurement is above measurement
range.

7

Measurement 13 Quality
Bad

8

Measurement 13 Low Limit

9

Measurement 13 High Limit

10

Measurement 14 Quality
Bad

11

Measurement 14 Low Limit

12

Measurement 14 High Limit

13

Measurement 15 Quality
Bad

14

Measurement 15 Low Limit

15

Measurement 15 High Limit

12 English