Modbus modbus register map – Hach-Lange SC 200 Modbus Module User Manual User Manual

Page 13

Advertising
background image

Modbus modbus register map

Group name

Tag name

Register # Data type

Length R/W Discrete

Range

Min/Max range Description

Setup

Baud rate

40001

Unsigned integer 1

R/W 0/1/2/3/4

Baud rate selection (0=9600; 1=19200;
2=38400; 3=57600; 4=115200)

Setup

Modbus mode

40002

Unsigned integer 1

R/W 0/1

Modbus mode (0=RTU; 1=ASCII)

Setup

Data order

40003

Unsigned integer 1

R/W 0/1

Register Data order (0=Little Endian
register order; 1=Big Endian register
order)

Setup

Parity

40004

Unsigned integer 1

R/W 2/0/1

Modbus parity (0=Even; 1=Odd;
2=None)

Setup

Stop bits

40005

Unsigned integer 1

R/W 1/2

Number of stop bits (1 or 2)

Setup/Addresses

Network card address

40006

Unsigned integer 1

R/W

0/246

Modbus address for the Modbus card
(1 to 246)

Setup

Modbus card name

40007

String

8

R/W

Location string of the network card

Network Timing

READ TIMEOUT

40015

Unsigned integer 1

R/W

1000/30000

Register read timeout setting (ms)

Network Timing

REG WRITE TMO

40016

Unsigned integer 1

R/W

3000/30000

Register write timeout setting (ms)

Network Timing

FILE WRITE TMO

40017

Unsigned integer 1

R/W

5000/30000

File write timeout setting (ms)

Network Timing

FILE PREP TMO

40018

Unsigned integer 1

6000/30000

File write timeout setting (ms)

Setup/Addresses

Device address

40019

Unsigned integer 1

R/W

0/246

The selected Modbus address of the
device (1 to 246)

Setup/Addresses

Select device

40020

Unsigned integer 1

R/W

0/30

Select the device to view/set Modbus
address (1 to 30)

Diagnostics

FUNCTION CODE

40021

Unsigned integer 1

R/W

0/65535

Function code used in the menu
system

Diagnostics

NEXT STATE

40022

Unsigned integer 1

R

0/65535

Next state value used in the menu
system

Diagnostics

INT BAUD RATE

40023

Unsigned integer 1

R

0/1/2/3/4

Baud rate selection (0=9600; 1=19200;
2=38400; 3=57600; 4=115200)

English 13