Figure 10 Jumper settings

1 Input 1 configuration jumpers

3 Input 3 configuration jumpers

5 Jumpers positioned to the right

for voltage inputs

2 Input 2 configuration jumpers

4 Jumpers positioned to the left for

switch inputs

1. Open the controller cover.
2. Feed the wires through the cable gland.
3. Adjust the wire as necessary and tighten the cable gland.
4. The jumpers are positioned immediately behind the connector. Remove the connector for

improved access to the jumpers and configure the jumper settings according to the type of input
as shown in

Figure 10

.

5. Close the controller cover and tighten the cover screws.
6. Configure inputs in the controller.

Note: In switch input mode the controller supplies 12 volts to the switch and is not isolated from the controller. In
voltage input mode the inputs are isolated from the controller (user input voltage from 0 to 30 volts).

Connect the optional digital communication output

The manufacturer supports Modbus RS485, Modbus RS232 and Profibus DPV1 communication
protocols. The optional digital output module is installed in the location indicated by item 4 in

Figure 7

on page 12. Refer to the manual supplied with the network module for more details.
For information about Modbus registers, refer to

http://www.hach-lange.com

or

http://www.hach.com

.

User interface and navigation

User interface

The keypad has four menu keys and four directional keys as shown in

Figure 11

.