Basic user manual, Additional information, Specifications – Hach-Lange POLYMETRON 9500 Basic User Manual User Manual

Page 3: General information, English

Advertising
background image

Additional information

Additional information is available on the manufacturer's website.

Specifications

Specifications are subject to change without notice.

Specification

Details

Component description

Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.

Operating temperature

-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W

Storage temperature

-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing

Enclosure

NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish

European standards

EN 61326-1: 2006; EN 61010-1:2010

ETL approved

File 65454

Power requirements

AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W
sensor/network module load, 100 VA with 28 W sensor/network module load
(optional Modbus RS232/RS485 or Profibus DPV1 network connection).

24 VDC powered controller: 24 VDC—15%, + 20%; Power 15 W with 7 W
sensor/network module load, 40 W with 28 W sensor/network module load (optional
Modbus RS232/RS485 or Profibus DPV1 network connection).

Altitude requirements

Standard 2000 m (6562 ft) ASL (Above Sea Level)

Pollution degree /
Installation category

Pollution Degree 2; Installation Category II

Outputs

Two analog (0-20 mA or 4-20 mA) outputs. Each analog output can be set to
0-20 mA or 4-20 mA, and can be assigned to represent a measured parameter such
as pH, temperature, flow or calculated values. Optional 3 additional analog outputs.
Secure Digital Memory card for use in data logging and software updates.

Relays

Four SPDT, user-configured contacts, rated 5A 250 VAC (resistive). Contacts are
rated 250 VAC, 5 Amp resistive maximum for the AC powered controller and 24 VDC,
5A resistive maximum for the DC powered controller. Relays are designed for
connection to AC Mains circuits (i.e., whenever the controller is operated with 115 -
240 VAC power) or DC circuits (i.e., whenever the controller is operated with 24 VDC
power).

Dimensions

½ DIN—144 x 144 x 180.9 mm (5.7 x 5.7 x 7.12 in.)

Weight

1.7 kg (3.75 lb)

Digital communication

Optional Modbus RS485/RS232 or Profibus DPV1 network connection for data
transmission

Data logging

Secure Digital Card or special RS232 cable connector for data logging and
performing software updates

Warranty

2 years

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.

English 3

Advertising