Introduction

3.1.3 Sensor recharge kit

A recharge kit (as illustrated in

Figure 4

below) should have been ordered with the sensor as

this will be required to initially make the sensor operational. It is also required for sensor
cleaning and membrane replacement procedures.

Note:

The recharge kit for oxygen has a blue sticker on the front of the box.

The kit contains:

four recharge cartridges with pre-mounted membrane and electrolyte. The type of
membrane mounted in the cartridge will be specific to the kit ordered

two anode cleaning tools

a set of five cotton washers

a set of five silicone discs (available in all 2956A-XXX recharge kits only)

a set of replacement O-rings

a set of Dacron

®

mesh patches.

The black anode cleaning tool (

Figure 5

) is used to clean the anode of any deposits or residue

that may have formed. It is doubled-ended so it can be used for two membrane replacement
processes, each end being used once.

Figure 4 Sensor recharge kit

Figure 5 Anode cleaning tool