Installation

4.2 Sensor installation

4.2.1 Sensor positioning information

Unless the sensor is part of the ORBISPHERE equipment that includes it, the sensor must be
installed in an ORBISPHERE socket or flow chamber, that allows contact with the sample fluid
to be analyzed.

The sensor and measuring instrument are connected by a cable and two 10-pin connectors.
The standard sensor cable length is 3 meters though extension cables of up to 1000 meters are
available. However, smart sensor technology is only available with distances of up to a
maximum of 750 meters.

Note:

If the model 28117 pressure sensor is used, the maximum cable length is 50 meters.

Ensure that the sensor will be mounted:

perpendicular to the pipe

horizontal

on a horizontal pipe section (or on flow-ascending vertical pipe)

min. 15 meters away from pump's discharge side

in a place where the sample flow is stable and rapid, and as far as possible from:

valves

pipe bends

the suction side of any pumps

a CO

2

injection system or similar

Note:

There may be situations where not all the above conditions can be met. If this is the case, or you

have any concerns, please consult your Hach Lange representative to appraise the situation and define the
best applicable solution.