Maintenance and troubleshooting – Hach-Lange ORBISPHERE A1100 Family User Manual

Page 38

Advertising
background image

36

Maintenance and Troubleshooting

Note:

During this step, it is very important to ensure

your finger does not come into contact with the
cathode (golden surface) as it could leave greasy
deposits on the surface.

31.

Take a silicone disc from the recharge kit,
hold it between thumb and forefinger and
position it on top of the anode.

Note:

As in the previous step, it is very important to

ensure your finger does not come into contact with the
cathode (golden surface) as it could leave greasy
deposits on the surface.

32.

Take a new cotton washer from the recharge
kit. Hold it between thumb and forefinger and
position it on top of the silicone disc.

33.

Place the recharge cartridge container on a
flat work surface and, keeping the container
upright to avoid spilling any of the electrolyte
inside, carefully unscrew the top.

34.

Remove the packing component from the
center of the cartridge, and make sure that
the O-ring remains in place on top of the
cartridge. If it comes away then replace it
before continuing.

35.

If there are any visible bubbles in the
electrolyte, remove them using a stirring
motion with the packing component.

Advertising