Additional display formats, Graphical display, Additional display formats graphical display – Hach-Lange POLYMETRON 9526 User Manual

Page 18

Advertising
background image

Figure 7 Example of Main Measurement screen

1 Home screen icon

7 Warning status bar

2 Sensor name

8 Date

3 SD Memory card icon

9 Analog output values

4 Relay status indicator

10 Time

5 Measurement value

11 Progress bar

6 Measurement unit

12 Measurement parameter

Table 1 Icon descriptions

Icon

Description

Home screen

The icon may vary depending on the screen or menu being displayed. For example, if an SD
card is installed, an SD card icon appears here when the user is in the SD Card Setup menu.

SD memory
card

This icon appears only if an SD card is in the reader slot. When a user is in the SD Card Setup
menu, this icon appears in the upper left corner.

Warning

A warning icon consists of an exclamation point within a triangle. Warning icons appear on the
right of the main display below the measurement value. Push the ENTER key then select the
device to view any problems associated with that device. The warning icon will no longer be
displayed once all problems have been corrected or acknowledged.

Error

An error icon consists of an exclamation point within a circle. When an error occurs, the error
icon and the measurement screen flash alternately in the main display. To view errors, push the
MENU key and select Diagnostics. Then select the device to view any problems associated
with that device.

Additional display formats

• From the Main Measurement screen push the UP and DOWN arrow keys to switch between

measurement parameters

• From the Main Measurement screen push the RIGHT arrow key to switch to a split display of up to

4 measurement parameters. Push the RIGHT arrow key to include additional measurements. Push
the LEFT arrow key as needed to return to the Main Measurement screen

• From the Main Measurement screen push the LEFT arrow key to switch to the graphical display

(see

Graphical display

on page 16 to define the parameters). Push the UP and DOWN arrow

keys to switch measurement graphs

Graphical display

The graph shows concentration and temperature measurements for each channel in use. The graph
supplies easy monitoring of trends and shows changes in the process.