Error list, Replacement parts and accessories – Hach-Lange POLYMETRON 9526 User Manual

Page 30

Advertising
background image

Table 3 Warning list for conductivity sensors (continued)

Warning

Description

Resolution

WRNG BLT-IN TEMP TABLE

The measured temperature is
outside of the range of the built-in
temperature compensation table

Make sure that the temperature
compensation is configured
correctly.

WRNG BLT-IN CONC TABLE

The concentration measurement is
outside of the range of the built-in
concentration table

Make sure that the concentration
measurement is configured for the
correct chemical and range.

Error list

Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in

Table 4

.

Table 4 Error list for conductivity sensors

Error

Description

Resolution

ADC FAILURE

The analog to digital conversion
failed

Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.

SENSOR MISSING

The sensor is missing or
disconnected

Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.

SENS OUT RANGE

The sensor signal is outside of the
accepted limits (2 S/cm)

Make sure that the display format is
set for the correct measurement
range.

Replacement parts and accessories

Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.

Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Replacement parts and accessories

Description

Item no.

Kit of 3 protection caps for instrument front panel connectors

09126=A=8010

Kit of 2 adapters D6/8 to DN4/6

09126=A=8020

Kit of 2 black protection caps for sample input and output connectors

09126=A=8030

Power supply connector

350=500=004

Tool for disconnecting in/out sampling tubes

578=507=602

DN8 semi-rigid PTFE tubing (per meter)

590=060=080

DN8 semi-rigid PE tubing (per meter)

151400,22387

Cable for the 4-20 mA output (5 meters)

08319=A=0005

Cable for the 4-20 mA output (10 meters)

08319=A=0010