Warning list – Hach-Lange POLYMETRON 9526 User Manual

Page 29

Advertising
background image

1. Push the menu key and select DIAGNOSTICS.
2. Select the device (controller, sensor, network card) with the warning or error and push enter.
3. Select the warning, error or event list and push enter.
4. Select YES and push enter to acknowledge the warning.

Note: Errors cannot be acknowledged.

Warning list

A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in

Table 3

.

Table 3 Warning list for conductivity sensors

Warning

Description

Resolution

MEAS TOO HIGH

The measured value is > 2 S/cm,
1,000,000 ppm, 200% or 20,000 ppt

Make sure that the display format is
set for the correct measurement
range

MEAS TOO LOW

The measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt

Make sure that the sensor is
configured for the correct cell
constant.

ZERO TOO HIGH

The zero calibration value is too
high

Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.

ZERO TOO LOW

The zero calibration value is too low

TEMP TOO HIGH

The measured temperature is >
200 °C

Make sure that the sensor is
configured for the correct
temperature element.

TEMP TOO LOW

The measured temperature is <
-20 °C

CAL OVERDUE

The Cal Reminder time has expired

Calibrate the sensor.

NOT CALIBRATED

The sensor has not been calibrated

Calibrate the sensor.

REPLACE SENSOR

The sensor has been in operation >
365 days

Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to

Sensor

diagnostic and test menu

on page 26. If the calibration fails,
call technical support.

CAL IN PROGRESS

A calibration was started but not
completed

Return to calibration.

OUTPUTS ON HOLD

During calibration, the outputs were
set to hold for a selected time.

The outputs will become active after
the selected time period.

WRONG LINEAR TC

The user-defined linear temperature
compensation is out of range

The value must be between 0 and
4%/°C; 0 to 200 °C.

WRONG TC TABLE

The user-defined temperature
compensation table is out of range

The temperature is above or below
the temperature range defined by
the table.

WRNG USER CONC TABLE

The concentration measurement is
outside of the range of the user
table

Make sure that the user table is set
for the correct measurement range.