Additional information

Additional information is available on the manufacturer's website.

Specifications

Specifications are subject to change without notice.

Analyzer

Specification

Details

Dimensions

Height: 450 mm; Width: 250 mm; Depth: 460 mm

Weight

7 kg (15.4 lb)

Casing protection

IP 65 / NEMA4X

Power supply

Standard version: 100-240 VAC 50/60 Hz

Low voltage version: 13-30 VAC 50/60 Hz, 18-42 VDC

Consumption: 25 VA

Measurement category: I (overvoltage less than 1,500 V)

Sample flow rate

20 liters/hour minimum

Sample tubing

Sample inlet and outlet: Diameter 8 mm (or 5/16'') semi-rigid tubing. We recommend the
use of PE tubing if sample temperature is inferior to 70 °C, and PTFE if superior to 70 °C

Connections

Power supply: Use the connector provided in the drawer

Analog output: Use the recommended POLYMETRON cable

Ambient temperature

-20 to 60 °C (-4 to 140 °F)

Maximum temperature 100 °C (at atmospheric pressure)

Maximum pressure

10 bar at ambient temperature

Relative humidity

10—90%

Precision

Conductivity: ± 2% of the displayed value

Temperature: ± 0.2 °C

Measurement range

Conductivity: 0.01 μS/cm to 200 μS/cm

Resistivity: 100 MΩ.cm to 5 kΩ.cm

Temperature: -20 to 200 °C (-4 to 392 °F)

Display resolution

0.001 μS/cm or 0.1 MΩ.cm

Outputs

Analog output (temperature, conductivity/resistivity): 2 × 0/4-20 mA (linear, bilinear,
logarithmic) ± 0.1 mA

Alarms: 2 × thresholds or limits according to USP

Certifications

EN 61326-1: 2006; EN 61010-1: 2010

Sensor

Specification

Details

Sensor body material

Black PSU

Conductivity electrodes, internal and external

Stainless steel 316L

English 3