2 using the digital gateway, Installation, 1 wiring the sc sensor to the digital gateway – Hach-Lange 3400 sc User Manual

Page 16

Advertising
background image

12

Installation

Figure 9:

Quick-connect Fitting pin assignment

3.2 Using the Digital Gateway

The digital gateway is designed to provide a digital interface to the controller. The
non-sensor end is wired to the controller as shown in

"Connecting the Sensor to an sc

Controller" on page 11

.

3.2.1 Wiring the sc Sensor to the Digital Gateway

1.

Route the cable from the sensor through the strain relief in the digital gateway then
properly terminate the wire ends.

Note: Do not tighten the strain relief until the digital gateway is wired and the two halves are
threaded securely together.

2.

Insert the wires as shown in

Table 5: "Wiring the Digital Gateway"

and

Figure 10:

"Wiring and Assembling the Digital Gateway"

.

3.

Make sure the O-ring is properly installed between the two halves of the digital
gateway and thread the two halves together. Hand tighten.

4.

Tighten the strain relief to secure the sensor cable.

5.

Connect the digital gateway to the controller.

Number

Designation

Wire Color

1

+12 VDC

Brown

2

Circuit Common

Black

3

Data (+)

Blue

4

Data (–)

White

5

Shield

Shield (grey wire in existing quick-disconnect fitting)

6

Groove

1

2

3

4

5

6

Table 5: Wiring the Digital Gateway

Sensor (wire color)

Sensor Signal

Digital Gateway Sensor Wire connector

Clear

Shield

J1-1

Clear w/shrink wrap

Shield

J1-1

Red

Drive

J1-2

White

Temp –

J1-3

Blue

Temp +

J1-4

Black

Sense

J1-5

Advertising