7 adjusting the temperature, Operation – Hach-Lange 3400 sc User Manual

Page 25

Advertising
background image

21

Operation

4.6.3.1 Preparing Conductivity Reference Solutions

Use

Table 6: "Conductivity Reference Solutions" on page 21

to prepare a conductivity

reference solution with a value between 200 and 100,000 µS/cm. The value prepared
should be near the typical measured process value for best accuracy. Add the listed grams
of pure, dried NaCl to one liter of high-purity, deionized, CO

2

-free water at 25 °C to obtain

the stated conductivity.

4.7 Adjusting the Temperature

View or change the temperature using the steps below.

1.

Select the Main Menu.

2.

From the Main Menu, select SENSOR SETUP and confirm.

3.

Select the appropriate sensor if more than one sensor is attached and confirm.

4.

Select DIAG/TEST and confirm.

5.

Select TEMP ADJUST and confirm.

The temperature will be displayed.

6.

Edit the temperature and confirm.

Table 6: Conductivity Reference Solutions

Desired Solution Value

Grams NaCl to be added

µS/cm

mS/cm

ppm (NaCl)

1

100

0.10

50

0.05

200

0.20

100

0.10

500

0.50

250

0.25

1000

1.00

500

0.50

2000

2.00

1010

1.01

3000

3.00

1530

1.53

4000

4.00

2060

2.06

5000

5.00

2610

2.61

8000

8.00

4340

4.34

10000

10.00

5560

5.56

20000

20.00

11590

11.59

1

When using the ppm measuring scale for compounds other than NaCl, refer to the appropriate chemistry handbook for reference
solution for formulation.