Table 1: 3400sc-series Conductivity Probe Specifications

Components

Corrosion-resistant materials, fully-immersible probe with 10 m (30 ft) cable

Measuring Range (Conductivity)

See

Table 3: " Sensor Cell Constants and Measuring Ranges" on page 5

Measuring Range (

Resistivity)

See

Table 3: " Sensor Cell Constants and Measuring Ranges" on page 5

Measuring Range (

TDS)

See

Table 3: " Sensor Cell Constants and Measuring Ranges" on page 5

Measuring Range (Temperature)

–20.0 to 200.0 °C (–4.0 to 392.0 °F)

Operating Temperature/Humidity

–20 to 60 °C (–4 to 140 °F); 0–95% relative humidity, non-condensing

Storage Temperature/Humidity

–30 to 70 °C (–22 to 158 °F); 0–95% relative humidity, non-condensing

Response Time

90% of reading within 30 seconds of step change

Measurement Accuracy

±2% of reading

Temperature Accuracy

±0.1 °C

Repeatability

±0.5% of reading

Sensitivity

±0.5% of reading

Calibration/Verification

Comparison to standard

Sensor Interface

Modbus

Standard Probe Cable Length

Analog probe: 6 m (20 ft); Digital probe: 10 m (32.8 ft)

Probe Weight

0.3 to 0.4 kg (approximately one pound) dependent on probe type

Probe Dimensions

Dependent on probe type,
see

Figure 1: "Compression-style Sensor, 0.5-in. Diameter" on page 8

through

Figure 7: "Boiler/Condensate Sensor" on page 10

.