1 specifications – Hach-Lange POLYMETRON 8350_8351 User Manual User Manual

Page 6

Advertising
background image

2/6

1 SPECIFICATIONS


The 8350 pH and 8351 redox probes are dedicated for process measurements (fouling) and are
protected against mechanical shochs.

All these pH probes incorporate process-contact temperature sensors and impedance control.

8350.4

8350.5

8351

Glass

Junction

Tube of Pt 100

Platinum ring

PTFE Junction

Stainless steel

tube of Pt 100

Glass (flat)

PTFE Junction

Platinum ring




NOTE: A special model, the 8350.3, is compatible with all the current, standard accessories and

is designed for use in contact with hydrofluoric acid.



8350.4 8350.5 8351

Application:

Measuring pH at
high temperatures

Measuring pH in a
fouling environment

Measuring redox

Junction type

Flat, PTFE

Flat, PTFE

Flat, PTFE

Body material

PPS

CPVC

PPS

Measurement range

0…14 pH

0…12 pH

± 1500 mV

Maximum temperature (°C)

110

80

110

Maximum pressure

10 bar

10 bar

10 bar

Drift (/ week)

< 3 mV

< 3 mV

< 3 mV

Offset @ pH7

± 20 mV

± 20 mV

-

Slope

56…61 mV / pH

56…61 mV / pH

-

Reference impedance @ 25 °C < 50 kohm

< 50 kohm

< 50 kohm

Glass impedance @ 25 °C

150 … 500 Mohm

50…250 Mohm

Platinum

Advertising