Hach-Lange POLYMETRON 8350_8351 User Manual User Manual

Page 8

Advertising
background image

4/6

VIA THE 08350=A=8000 PRE-AMPLIFIER UNIT


It is impossible to connect the liquid earth in this case.

Ext.

Shielding

white

pink

yellow

brown

grey

green

Input (pH/rH)

- V

+ V

Ref. (Reference)

Temp + (Pt 100)

Temp - (Pt 100)

EMC

protection

EMC

protection

Input (pH/rH)

Ref. (Reference)

Temp + (Pt 100)

Temp - (Pt 100)

white

red

black

Clear

Ext.

Shielding

2666 cable

08350=A=8000 Pre-amplifier

Probe




USING THE 08350=A=8500 JUNCTION BOX

white

red

black

Clear

Probe

Temp + (Pt 100)

Temp - (Pt 100)

Ref. (Reference)

Ext.

Shielding

white

Schielding

Input (pH/rH)

EMC

protection

Core

yellow / green

violet

Liquid earth (contact)

EMC

protection

Liquid earth (contact)

Temp + (Pt 100)

Temp - (Pt 100)

Ref. (Reference)

Input (pH/rH)

Ext.

Shielding

blue

08350=A=8500 Junction box

2654

For pH probes only:
Temperature measurement.
-

Impedance measurement (the blue wire from the transmitter's "liquid earth"): programme the control frequency and the upper and lower
limits (refer to the instructions for the 9135 transmitter):

Glass impedance:

50 Mohm - 500 Mohm

Reference impedance:

1…100 kohm