2 INSTALLATION


The 8350 probes should be installed vertically for optimal results. They may, however, be installed
horizontally or upside down.
• Insertion mounting:

screw-in or flange fitting.

• Immersion

mounting: immersion

shafts (0.5m; 1m; 1.5m) with fixed, adjustable or clip-type

flanges.

The probes are fitted with 10 metres of cable as standard.
For longer cable lengths, up to 25m, use a straight junction box (08350=A=8500) and the 2654
cable (358048,00000).
For even longer cable lengths, use a pre-amplifier unit (08350=A=8000) and the 2666 cable
(370=506=025).
In this case it will not be possible to monitor the impedance.

3 CONNECTION TO THE 9135 TRANSMITTER

DIRECTLY TO THE 9135

External

shielding

EMC

protection

Input (measured pH or rH)

Ref. (reference)

Liquid earth (liquid contact)

white

red

clear

black

blue

Temp + (pt100)

Temp - (pt100)