Specifications

The 8350 pH and 8351 Redox probes are dedicated for process measurements. The pH probes
incorporate impedance control and a temperature sensor in contact with the sample.

8350.4

8350.5

8351

1 - Glass

2 - Tube containing PT100

3 - Platinum ring

4 - PTFE Junction

A special model (8350.3) is compatible with all the standard accessories and is designed for use in
hydrofluoric acid.

8350.4

8350.5

8351

Application

Measuring pH at high

temperatures

Measuring pH in a

waste water

environment

Measuring redox

Junction type

Flat, PTFE

Flat, PTFE

Flat, PTFE

Body material

PPS

CPVC

PPS

Measurement range

0-14 pH

0-12 pH

± 1500 mV

Maximum temperature (°C)

110

80

110

Maximum pressure

10 bar

10 bar

10 bar

Drift (/ week)

< 3 mV

< 3 mV

< 3 mV

Offset @ pH7

± 20 mV

± 20 mV

N/A

Slope

56-61 mV/pH

56-61 mV/pH

N/A

Reference impedance @ 25°C

< 50 Kohm

< 50 Kohm

< 50 Kohm

Glass impedance @ 25°C

150-500 Mohm

150-500 Mohm

Platinum

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.

Safety information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention
to all danger and caution statements. Failure to do so could result in serious injury to the operator or
damage to the equipment.