Using the junction box, Calibration – Hach-Lange POLYMETRON 8350_8351 Basic User Manual User Manual

Page 6

Advertising
background image

1

2666 cable

9

-V - grey

2

08350=A=8000 pre-amplifier

10 Input (measured pH or redox) - green

3

Probe

11 Temp+ (PT100) - white

4

External shielding - EMC protection

12 Temp- (PT100) - red

5

Temp+ (PT100) - white

13 Ref. (reference) - black

6

Temp- (PT100) - pink

14 Input (measured pH or redox) - transparent

7

Ref. (reference) - yellow

15 External shielding - EMC protection

8

+V - brown

Using the junction box

1

2654 cable

9

External shielding - EMC protection

2

08350=A=8500 junction box

10 External shielding - EMC protection

3

Probe

11 Input (measured pH or redox) - transparent

4

Input (measured pH or redox) - core

12 Ref. (reference) - black

5

Ref. (reference) - shielding

13 Temp+ (PT100) - white

6

Temp+ (PT100) - white

14 Temp- (PT100) - red

7

Temp- (PT100) - violet

15 Liquid earth (liquid contact) - blue

8

Liquid earth (liquid contact) - yellow/green

For pH probes only:

• Temperature measurement
• Impedance measurement (blue wire from the transmitter):

• Program the control frequency upper and lower limits (refer to the 9135 operator manual):

• Glass impedance: 50 Mohm to 500 Mohm
• Reference impedance: 1 to 100 kohm

Calibration

Before calibrating for the first time:

1. Shake the probe gently to remove any air bubbles from underneath the glass bulb.
2. Remove the storage cap and rinse the probe with demineralized water.

6 English