Connection to the 9135 transmitter, Directly to the 9135, Using the pre-amplifier unit – Hach-Lange POLYMETRON 8350_8351 Basic User Manual User Manual

Page 5

Advertising
background image

The probes are fitted with 10 metres of cable as standard.
For longer cable lengths (up to 25 meters) use a junction box (08350=A=8500) and the 2654 cable
(358048,00000).
For cable lengths greater than 25 meters, use a pre-amplifier unit (08350=A=8000) and the 2666 cable
(370=506=025). In this case it will not be possible to monitor the impedance.

Connection to the 9135 transmitter

Directly to the 9135

1 pH module

5 Input (measured pH or redox) - transparent

2 Probe

6 Ref. (reference) - black

3 Temp+ (PT100) - white

7 Liquid earth (liquid contact) - blue

4 Temp- (PT100) - red

8 External shielding - EMC protection

Using the pre-amplifier unit
With this scenario it is impossible to connect the liquid earth.

English 5