Storage, Troubleshooting, If a calibration error occurs, refer to – Hach-Lange LDO10130 User Manual

Page 10: If a measurement error occurs, refer to

Advertising
background image

Storage

Dry storage—the manufacturer recommends that the probe is stored dry when the
probe is used for measurements of short duration (less than 6 hours).

Wet storage—the probe must be stored wet when the probe is used for monitoring
periods longer than 6 hours.

Dry storage

Note: Rugged probes may be stored dry with the shroud installed if the storage container is

sufficiently large.

1. Rinse the probe with deionized water. Blot dry with a lint-free cloth.

Note: The probe must be conditioned again in tap water for at least 30 minutes prior to use.

2. If a rugged probe, remove the shroud (optional). Refer to

Remove the shroud

on page 9.

Wet storage

Note: The need for recalibration is minimized if the probe cap is kept wet.

1. Put the probe in tap water.
2. During the initial 72 hours in tap water, calibrate the probe once every 8 hours.

Note: After 72 hours of storage in tap water, the probe cap will reach a fully hydrated state.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the software,
refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe requires
service

Probe not connected
correctly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

To download the most current version of the software,
refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods
stored on the probe

Continue to let the probe connect. Do not disconnect the
probe.

Damaged probe

Make sure there is connectivity with another probe or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.