Change calibration options

1. Make sure that a probe is connected to the meter.
2. Push and select LDO101 Settings.
3. Select Modify Current Settings.
4. Select Calibration Options and update the settings:

Option

Description

Calibration

User—100% (water-saturated air (100%) calibration)

User—100% with 0 (water-saturated air (100%) calibration with 0 point)

User—mg/L (calibration with a specified dissolved oxygen concentration
(mg/L) solution)

User—mg/L with 0 (calibration with a specified dissolved oxygen
concentration (mg/L) solution with 0 point )

Factory (calibration with the default LDO calibration)

Standard
Value

When Calibration is set to mg/L or mg/L with 0, sets the concentration of the
solution that will be used for calibration—2.00 to 20.00 mg/L
(default=7.00 mg/L)

5. Select Calibration Reminder and update the settings:

Option

Description

Reminder
Repeat

Meter will make an audible sound when calibration is due and repeat the
sound at selected interval—Off, 8 h, 12 h, 1 d, 2 d, 5 d or 7 d.

Expires

Calibration expires after the selected time—Immediately, Reminder +
30 min, Reminder + 1 h, Reminder + 2 h or Continue Reading.
Note: The meter cannot be used to read samples after calibration has

expired unless Continue Reading is selected.

6. If prompted, enter a name for the new method settings. Additional changes made to

the settings of an existing method are automatically saved with the same method
name.

7. Push EXIT until the meter goes back to the measurement mode.

Maintenance

Clean the probe

Keep the probe cap free of deposits for the best measurements.

Note: Do not touch the black colored substrate of the probe cap. Do not use alcohol or other

organic solvents to clean the black colored substrate of the probe cap. These solvents cause

damage to the probe cap.

1. Remove the shroud (refer to

Remove the shroud

on page 9).

2. Gently clean the probe cap with a mild detergent, water and a soft cloth or cotton

swab. Do not remove the black colored substrate from the probe cap. Do not scrub
the probe cap or lens.

3. If water is present between the probe cap and lens:

a. Remove the probe cap.
b. Blot dry the probe cap and lens with a soft dry cloth.
c. Install the probe cap.

4. Install the shroud (refer to

Install the shroud

on page 9).