Message or symptom

Possible cause

Action

Calibration failed: outside of
acceptance criteria/Temperature
out of range/Offset out of limits

Water Saturated air
equilibration not reached

Allow longer equilibration time.

Probe cap loose, scratched,
or damaged

Change the location of the probe cap or replace the
probe cap.

Temperature and/or pressure
sensor error

Make sure that the temperature and pressure sensors
are both reading correctly and within range.

1

Damaged probe

Make sure that the blue and red LEDs are both
illuminated on the probe. If not, replace the probe or
contact a Technical Support Representative.

1

The pressure as measured by the probe is what is referred to as atmospheric pressure and is not corrected to sea level.
Weather station pressures are reported at sea level and commonly referred to as mean sea level pressure. As a result the
probe will not read the same as most household or professional barometers or weather station reports (which are
compensated) unless reported at sea level. In order to compare the pressure results obtained from the probe barometer
and these compensated barometers, it is necessary to first compensate the pressure reported by the probes
mathematically.

HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. (970) 669-3050
(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
[email protected]
www.hach.com

HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
[email protected]
www.hach-lange.de

HACH LANGE Sàrl
6, route de Compois
1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499

©

Hach Company/Hach Lange GmbH, 2010, 2013. All rights reserved. Printed in U.S.A.

05/2013, Edition 2

Advertising
This manual is related to the following products: