Preparation for use – Hach-Lange LDO10130 User Manual

Page 2

Advertising
background image

probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory
use. The probe measures the dissolved oxygen concentration in wastewater, drinking
water and general applications.

Note: For BOD applications, use the LBOD10101 probe which has LDO technology and a stirrer.

Figure 1 Probe overview

1 LDO probe cap

5 Pressure sensor module

2 Rugged probe (5, 10, 15 or 30 meter cable)

6 iButton

®

compartment

1

3 Standard probe (1 or 3 meter cable)

7 Pressure sensor module cap

4 Probe cable

1

iButton is a registered trademark of Maxim Integrated Products, Inc.

Preparation for use

Note: Do not touch the probe cap with a hand, fingers or any surface that can scratch the cap.

Prepare the probe for use before calibration or sample measurement.

1. Make sure that the probe cap and iButton are installed correctly. The iButton label

should be up.

2. Make sure that the probe cap and iButton have the same lot code.
3. If a rugged probe, make sure that the shroud is installed before field use (refer to

Install the shroud

on page 9).

Note: Damage to the sensing elements can occur if the shroud is not installed during field use.

Damage under these conditions is not covered by the product warranty.

4. Rinse the probe cap with deionized water. Blot dry with a lint-free cloth.
5. If dissolved oxygen monitoring periods are longer than 6 hours, condition the probe

cap for 72 hours. Calibrate the probe once every 8 hours.

Note: After 72 hours, the probe cap will reach a fully hydrated state.

2